Week 8 - Testing Plethysmograph circuit

Saturday, 29 September 2012

As salam..

This post is continued from the previous post.
The research on the Plethysmograph is helped
for me to test the circuit that shown in the schematic
below:
pulse plethysmograph circuit
Pulse Plethysmograph Amplifier Circuit



Sensor
The sensor consists of a light source and photodetector.
Whereby the light is shone through the tissues and 
variation in blood volume alters the amount of light falling 
on the detector. The light source and detector can be 
mounted side by side to look at changes in reflected light 
or on either side of a finger or earlobe to detect changed 
in transmitted light. The infrared filter of the phototransistor 
reduces interference from fluorescent lights, which have a 
large AC component in their output.

Amplifier
The amplifier uses an LM358 dual op-amp to provide two 
identical broadly-tuned band pass stages with gains of 100.
The circuit runs from a single 6V battery and the output 
zero is offset by about 1V by referring everything to an 
internal common line at a voltage set by a pair of 
forward-biased silicon diodes. This is convenient for 
interfaces with a 0-5V input.

Components are not critical but the two 2.2uF capacitors 
must be able to stand some reverse bias so they should be 
non-polarized. To get the expected result, the potentiometer 
is adjusted, so that the output is about 2V peak. The trace 
should look like this:
calculating heart rate experiment results
Heart rate trace

calculating heart rate results
Pulse Plethysmograph results during deep breathing
Actually this experiment's result is achieved by using the 
PicoLog software. From this software the expectation 
waveform will display clearly as required. 

So, I'm just trying the circuit by using the oscilloscope 
in lab. The sensor I changed to LDR and Red LED.
The result I obtained is shown as the pictures below:

When the sensor not touched
When the sensor is touched
The first waveform is produced is quite same with
the expected result just had the noise.
The problem I had is, when I'm connected to 
ground, the waveform is not displayed at above.
It was displayed like too much of noise and 
no pattern of waveform can be seen. When I'm 
not connected the ground the result is displayed 
at above.

The potentiometer I'm already adjusted but the 
waveform is still not give any pattern if I'm 
connected to the ground. If I'm disconnected the 
ground the waveform is displayed. 

The sensor I'm already change to the 
IR transmitter and receiver. But the result still the 
same.
The circuit was constructed

0 comments:

Post a Comment

Note: only a member of this blog may post a comment.